Title: | Reads Mass Spectrometry Data in Bruker *flex Format |
---|---|
Description: | Reads data files acquired by Bruker Daltonics' matrix-assisted laser desorption/ionization-time-of-flight mass spectrometer of the *flex series. |
Authors: | Sebastian Gibb [aut, cre] , Samuel Granjeaud [ctb], Alan Race [ctb] |
Maintainer: | Sebastian Gibb <[email protected]> |
License: | GPL (>= 3) |
Version: | 1.9.3 |
Built: | 2024-11-13 06:14:08 UTC |
Source: | https://github.com/sgibb/readbrukerflexdata |
The readBrukerFlexData package reads data files acquired by MALDI-TOF MS on
Bruker Daltonics machines of the *flex series. (autoflex,
microflex, ultraflex).
The package was developed without any knowledge nor even support by
Bruker Daltonics.
All trademarks are owned by or licensed to Bruker Daltonics.
Sebastian Gibb [email protected]
https://github.com/sgibb/readBrukerFlexData
Useful links:
Report bugs at https://github.com/sgibb/readBrukerFlexData/issues/
This dataset was generated by Bruker Daltonics CompassXport and imported by
readMzXmlFile
to R. It is only needed for
comparison between Bruker Daltonics' HPC and
.hpc
.
A list containing a mass and an intensity vector.
Examples/hpc/mzXML/hpc.mzXML
This function leads recursively all mass spectrometry data in Bruker Daltonics XMASS format in a specified directory.
readBrukerFlexDir( brukerFlexDir, removeCalibrationScans = TRUE, removeMetaData = FALSE, useHpc = TRUE, useSpectraNames = TRUE, filterZeroIntensities = FALSE, verbose = FALSE )
readBrukerFlexDir( brukerFlexDir, removeCalibrationScans = TRUE, removeMetaData = FALSE, useHpc = TRUE, useSpectraNames = TRUE, filterZeroIntensities = FALSE, verbose = FALSE )
brukerFlexDir |
|
removeCalibrationScans |
|
removeMetaData |
|
useHpc |
|
useSpectraNames |
|
filterZeroIntensities |
|
verbose |
|
See readBrukerFlexFile
.
A list
of spectra.
[[1]]$spectrum$mass
: A vector of calculated mass.
[[1]]$spectrum$intensity
: A vector of intensity values.
[[1]]$metaData
: A list of metaData depending on read spectrum.
importBrukerFlex
,
readBrukerFlexFile
,
.hpc
## load library library("readBrukerFlexData") ## get examples directory exampleDirectory <- system.file("Examples", package="readBrukerFlexData") ## read example spectra spec <- readBrukerFlexDir(file.path(exampleDirectory, "2010_05_19_Gibb_C8_A1")) ## plot spectra plot(spec[[1]]$spectrum$mass, spec[[1]]$spectrum$intensity, type="n") l <- length(spec) legendStr <- character(l) for (i in seq(along=spec)) { lines(spec[[i]]$spectrum$mass, spec[[i]]$spectrum$intensity, type="l", col=rainbow(l)[i]) legendStr[i] <- spec[[i]]$metaData$fullName } ## draw legend legend(x="topright", legend=legendStr, col=rainbow(l), lwd=1)
## load library library("readBrukerFlexData") ## get examples directory exampleDirectory <- system.file("Examples", package="readBrukerFlexData") ## read example spectra spec <- readBrukerFlexDir(file.path(exampleDirectory, "2010_05_19_Gibb_C8_A1")) ## plot spectra plot(spec[[1]]$spectrum$mass, spec[[1]]$spectrum$intensity, type="n") l <- length(spec) legendStr <- character(l) for (i in seq(along=spec)) { lines(spec[[i]]$spectrum$mass, spec[[i]]$spectrum$intensity, type="l", col=rainbow(l)[i]) legendStr[i] <- spec[[i]]$metaData$fullName } ## draw legend legend(x="topright", legend=legendStr, col=rainbow(l), lwd=1)
This function reads mass spectrometry data in Bruker Daltonics XMASS format used by Bruker Daltonics mass spectrometer of *flex series (autoflex, microflex, ultraflex).
readBrukerFlexFile( fidFile, removeMetaData = FALSE, useHpc = TRUE, filterZeroIntensities = FALSE, keepNegativeIntensities = FALSE, verbose = FALSE )
readBrukerFlexFile( fidFile, removeMetaData = FALSE, useHpc = TRUE, filterZeroIntensities = FALSE, keepNegativeIntensities = FALSE, verbose = FALSE )
fidFile |
|
removeMetaData |
|
useHpc |
|
filterZeroIntensities |
|
keepNegativeIntensities |
|
verbose |
|
readBrukerFlexFile
has to import the following data to calculating
mass from acqu file:
acqu-value | becomes metaData | description |
$BYTORDA | metaData$byteOrder | endianness of fid file |
$TD | metaData$number | total number of measured time periods |
$DELAY | metaData$timeDelay | first measured intensity after metaData$timeDelay ns |
$DW | metaData$timeDelta | ns between measured time periods |
$ML1 | metaData$calibrationConstants[1] | mass calibration constant |
$ML2 | metaData$calibrationConstants[2] | mass calibration constant |
$ML3 | metaData$calibrationConstants[3] | mass calibration constant |
If High Precision Calibration (HPC) is used, readBrukerFlexFile
needs:
acqu-value | becomes metaData | description |
$HPClBHi | metaData$hpc$limits[“maxMass”] | upper mass threshold |
$HPClBLo | metaData$hpc$limits[“minMass”] | lower mass threshold |
$HPClOrd | metaData$hpc$order | polynomial order |
$HPClUse | metaData$hpc$use | maybe using of HPC? (seems to be always “yes” in our test data) |
$HPCStr | metaData$hpc$coefficients | polynomial coefficients in a string |
readBrukerFlexFile
tries also to import [optional]:
acqu-value | becomes metaData | description |
DATATYPE | metaData$dataType | e.g CONTINUOUS MASS SPECTRUM |
SPECTROMETER/DATASYSTEM | metaData$dataSystem | e.g. Bruker Flex Series |
.SPECTROMETER TYPE | metaData$spectrometerType | e.g. TOF |
.INLET | metaData$inlet | DIRECT |
.IONIZATION MODE | metaData$ionizationMode | e.g. LD+ |
$DATE | metaData$date | same as $AQ_DATE but often only "0" |
$ACQMETH | metaData$acquisitionMethod | path to method file |
$AQ_DATE | metaData$acquisitionDate | acquisition date |
$AQ_mod | metaData$acquisitionMode | acquisition mode |
$AQOP_m | metaData$acquisitionOperatorMode, metaData$tofMode | LINEAR / REFLECTOR |
$ATTEN | metaData$laserAttenuation | laser beam attenuation |
$CMT[1:4] | metaData$comments | comments |
$DEFLON | metaData$deflection | deflection ON/OFF |
$DIGTYP | metaData$digitizerType | type of digitizer |
$DPCAL1 | metaData$deflectionPulserCal1 | deflection pulser cal 1 |
$DPMASS | metaData$deflectionPulserMass | deflection pulser mass |
$FCVer | metaData$flexControlVersion | Version of Bruker Daltonics FlexControl software |
$ID_raw | metaData$id | spectrum id |
$INSTRUM | metaData$instrument | e.g. AUTOFLEX |
$InstrID | metaData$instrumentId | ID of mass spectrometer |
$InstTyp | metaData$instrumentType | instrument type |
$Lift1 | metaData$lift[1] | LIFT constant? |
$Lift2 | metaData$lift[2] | LIFT constant? |
$Masserr | metaData$massError | initial mass error in ppm |
$NoSHOTS | metaData$laserShots | number of applied laser shots |
$PATCHNO | metaData$patch | sample postion on target |
$PATH | metaData$path | original file path (on Bruker *flex series controller PC) |
$REPHZ | metaData$laserRepetition | laser repetition rate in Hz |
$SPOTNO | metaData$spot | same as $PATCHNO
(in older files often empty, that's why readBrukerFlexFile
uses $PATHNO instead) |
$SPType | metaData$spectrumType | e.g. TOF |
$TgIDS | metaData$target$id | target ids |
$TgCount | metaData$target$count | number of measurements with this target |
$TgSer | metaData$target$serialNumber | target serial number |
$TgTyp | metaData$target$typeNumber | target type number |
$TLift | metaData$tlift | LIFT constant? |
import from file path:
value | becomes metaData | description |
full current path to fid file | metaData$file | path on local machine |
sample name | metaData$sampleName | - |
filterZeroIntensities
: Change default value is not recommended!
If TRUE
all intensities equal zero are removed.
This parameter exists only to be compatible to
Bruker Daltonics CompassXport's mzXML export function. For details see:
‘Release Notes for CompassXport 3.0.3’,
cap. 6 ‘Filtering of Zero Intensities’:
“Bruker Daltonics' Acquisition Software will compress Analysis raw
data. To save on operation time and to keep export file sizes small,
CompassXport 3.0.3 will filter out zero (0.0) intensities
when exporting to mzXML or mzData ...”
keepNegativeIntensities
: Change default value is
not recommended! If TRUE
negative intensity values are not
replaced by zero. This parameter exists only to be compatible to
Bruker Daltonics CompassXport.
A list
of spectra and metadata.
spectrum$mass
: A vector of calculated mass.
spectrum$tof
: A vector of time-of-flight data.
spectrum$intensity
: A vector of intensity values.
metaData
: A list of metaData depending on read spectrum.
https://github.com/sgibb/readBrukerFlexData/wiki,
importBrukerFlex
,
readBrukerFlexDir
,
.hpc
## load library library("readBrukerFlexData") ## get examples directory exampleDirectory <- system.file("Examples", package="readBrukerFlexData") ## read example spectrum spec <- readBrukerFlexFile(file.path(exampleDirectory, "2010_05_19_Gibb_C8_A1/0_A1/1/1SLin/fid")) ## print metaData print(spec$metaData) ## plot spectrum plot(spec$spectrum$mass, spec$spectrum$intensity, type="l", col="red")
## load library library("readBrukerFlexData") ## get examples directory exampleDirectory <- system.file("Examples", package="readBrukerFlexData") ## read example spectrum spec <- readBrukerFlexFile(file.path(exampleDirectory, "2010_05_19_Gibb_C8_A1/0_A1/1/1SLin/fid")) ## print metaData print(spec$metaData) ## plot spectrum plot(spec$spectrum$mass, spec$spectrum$intensity, type="l", col="red")